expanded uncertainty
**Expanded Uncertainty** ($U$) is the **combined standard uncertainty multiplied by a coverage factor to provide a confidence interval** — $U = k cdot u_c$, where $k$ is typically 2 (providing approximately 95% confidence) or 3 (approximately 99.7% confidence) that the true value lies within the stated interval.
**Expanded Uncertainty Details**
- **k = 2**: ~95% confidence level — the most common reporting convention.
- **k = 3**: ~99.7% confidence level — used for safety-critical or high-consequence measurements.
- **Reporting**: $Result = x pm U$ (k = 2) — standard format for reporting measurement results with uncertainty.
- **Student's t**: For small effective degrees of freedom, use $k = t_{95\%,
u_{eff}}$ from the t-distribution.
**Why It Matters**
- **Communication**: Expanded uncertainty communicates measurement quality in an intuitive way — "the true value is within ±U with 95% confidence."
- **Conformance**: Guard-banding uses expanded uncertainty to prevent accepting out-of-spec product — adjust limits by ±U.
- **Standard**: ISO 17025 accredited labs must report expanded uncertainty with measurement results.
**Expanded Uncertainty** is **the confidence interval** — combined uncertainty scaled by a coverage factor to provide a meaningful confidence statement about the measurement result.