Secondary-Ion Mass Spectrometer

SIMS and time-of-flight SIMS (ToF-SIMS) both measure composition by sputtering a sample's surface with a primary ion beam and analyzing the secondary ions that fly off — a destructive technique, by design, that reaches dopant and trace-element sensitivity no other analytical method in the fab's toolkit comes close to matching.

Trace-level or depth-resolved composition question
Dopant profile, trace contamination far below EDS/EPMA detection limits
Secondary ion mass spectrometry mode
Primary ion beam sputters surface
Removes a thin layer, ejecting secondary ions
Mass spectrometer analyzes ejected ions
Quadrupole (standard SIMS) or time-of-flight (ToF-SIMS)
↻ next depth layer
Depth-resolved composition profile, trace sensitivity
Parts-per-billion range for many elements

SIMS reaches trace sensitivity that EDS and EPMA fundamentally can't, because it directly counts individual sputtered ions rather than relying on a bulk X-ray signal that trace-level elements barely contribute to. A dopant present at parts-per-million or lower concentration generates far too weak an X-ray signal for EDS or EPMA to detect reliably against the background from the bulk material; SIMS instead physically removes and counts individual ions as the surface sputters away, giving it detection limits in the parts-per-billion range for many elements — which is exactly the sensitivity needed to measure an actual implanted dopant profile, something EDS and EPMA simply cannot do.

SIMS and ToF-SIMS — Sputter, Count, Repeat Deeper destructive by design, trading the sample for extreme trace sensitivity Depth profile — dopant concentration vs. depth implanted dopant profile resolved by depth SIMS vs ToF-SIMS analyzer quadrupole SIMS depth profiling, one mass at a time ToF-SIMS full mass spectrum per pulse, surface imaging too ToF-SIMS's parallel mass detection enables chemical imaging that scanning quadrupole SIMS can't do as well

ToF-SIMS adds surface chemical imaging capability that standard quadrupole SIMS wasn't built for, by capturing an entire mass spectrum from every single pulse instead of scanning through masses sequentially. Conventional SIMS typically tunes its detector to one mass at a time, which is efficient for depth profiling a known set of elements but awkward for building a chemical image across a surface; ToF-SIMS's time-of-flight detector captures essentially the full mass spectrum from each primary-ion pulse simultaneously, which supports both depth profiling and pixel-by-pixel chemical imaging of a surface — a dual capability standard SIMS doesn't offer as naturally.

SIMS and ToF-SIMS both accept destroying part of the sample as the price for reaching dopant and trace-contamination sensitivity nothing else in the analytical toolkit matches, and the choice between the two formats comes down to whether the job is a straightforward depth profile or needs the added chemical-imaging capability ToF-SIMS's parallel detection provides.